GL GONIO PHOTOMETER GLG A 50-1800

Type A goniophotometer for photometric measurements in the automotive industry

GL GONIO PHOTOMETER GLG A 50-1800 automotive light measurements

Goniophotometer will improve productivity and accelerate in-house product development

The continuous development of technology translates into an urgent need to speed up product introduction. This applies to every industry, including automotive lighting manufacturing. The series of tests that vehicle exterior lamps must undergo in the approval process is causing lighting manufacturers to choose to build their own laboratories to reduce the time to market for the final product.

Standard-compliant goniophotometer

GL GONIO PHOTOMETER GLG A 50-1800 has been designed in accordance with CIE 121-1996 and IESNA LM-75-01 standards governing photometric and colorimetric far-field measurement systems. It allows the determination of photometric characteristics in H, V coordinates for automotive lamps and headlights as well as all kinds of signaling equipment used, for example, in aviation.
The GL Optic goniophotometer is characterized by its high precision, robust design and user-friendly interface. In cooperation with a photometer, spectrometer and retroreflectometer, it forms a complete system for the automotive industry.

Compliance with international standards

GL Optic’s equipment is constantly updated to work in accordance with internationally available standards. The GLG A 50-1800 goniophotometer works in accordance with CIE 121-1996 and IESNA LM-75-01 standards, and the system enables photometric testing in accordance with UN/ECE instructions and FMVSS requirements based on SAE standards.

Reliability and accuracy

The GLG A 50-1800 goniophotometer is built with the highest quality mechatronic components from leading manufacturers. This ensures high accuracy of the entire system and repeatability of movements. The design of the goniophotometer allows the measurements of lamps weighing up to 50 kg and measuring up to 1800 mm length (optional up to 2600 mm).

Faster than the competition

The goniophotometric system working with the GL PHOTOMETER 3.0 LS + Flicker reduces measurement time by up to 5 times! The high sampling rate together with the high sensitivity of the sensor allows continuous On-fly measurement during smooth lamp movement. Using the flicker measurement function, it is possible to determine the light modulation characteristics.

GL GONIO PHOTOMETER GLG A 50-1800 Usage

Increased possibilities

Designers are racing to constantly improve and create unconventional lamps and headlamps. The development of LED technology also contributes to the creation of increasingly unusual solutions. In order to keep up with these developments, testing is necessary at the prototyping stage, preferably in-house in the R&D or QC.

High-accuracy goniophotometer

The GLG A 50-1800 goniophotometer can be used in both factory and accredited laboratories. It allows reliable data to be quickly obtained for large and smaller lamps. The design of the goniometer allows the testing of luminaires weighing up to 50 kg and with a diameter of 1800 mm (with an option of up to 2600 mm on special order).
The accuracy of the measurements is guaranteed by 3 motorized axes with absolute encoders. Extensive automation capabilities and intuitive software ensure a new level of productivity.

Faster lamp testing

GL PHOTOMETER 3.0 LS + Flicker is a dedicated device for use with a goniometer. It offers the possibility to perform On-fly measurements. During On-fly measurement, the goniometer arm move at a constant speed without stopping, while the photometer takes many thousands of measurements per second to accurately determine the photometric light curve for each plane.
In addition, the flicker measurement function allows the light modulation characteristics of indicators and emergency lights to be determined.

Color measurement also possible

The GL Optic goniophotometric system can be supplemented with a spectroradiometer to extend its functionality to include color measurement. The GL SPECTIS 1.3 LS is optimized for high-speed measurements of photometric and colorimetric parameters. It is ideal for basic applications and radiation measurements in the visible range.
GL SPECTIS 5.0 Touch can be used in advanced systems. It also enables spectral analysis outside the visible range (UV-NIR).

GL GONIO PHOTOMETER GLG A 50-1800 Features

The standard goniometric system includes:

  • Robust, accurate and programmable A-type goniometer with three motorized H, V and Z axes and mounting table, movable in x, y axes
  • Dedicated PC with GL SPECTROSOFT software
  • High-speed GL PHOTOMETER 3.0 LS + Flicker, enabling On-fly measurements and flicker characterization using PWM
  • Unique laser alignment system with mirrors to help with vertical and horizontal calibration of the system as well as faster photometric positioning of DUTs

Optional peripherals

  • GL SPECTIS 1.3 LS – spectroradiometer optimized for high-speed photometric and colorimetric measurements
  • GL SPECTIS 5.0 Touch – spectroradiometer for extended spectral analysis beyond the visible range from UV to NIR in various applications
  • Power supplies and meters – available power supplies and power sources allow full electrical characterization of the DUT. Advanced power meters are also available on request
  • TEC controller – for thermal stabilization during goniometric tests

GL GONIO PHOTOMETER GLG A 50-1800 Metrics

Photometric quantities

  • Light distribution
  • Luminous flux
  • Maximum brightness
  • IES and LDT files

Colorimetric quantities

  • Color temperature, CCT and Duv
  • Color rendering index, CRI (Ra)
  • MacAdam ellipses
  • Angular color uniformity

Other quantities

  • Luminous efficacy [lm/W]
  • Power factor cosφ
  • Temperature

GL GONIO PHOTOMETER GLG A 50-1800 Specifications

APPLICATION
Compliance testing of lamps and other lighting devices accor ding to UN/ECE and SAE /FMVSS Regulations.
This is a Type A system compliant with the CIE 121-1996 and IESNA LM-75-01 standards for far-field photometric and colorimetric measurement systems.
TECHNICAL DATA SHEET
CIE Goniometer typeFar Field Type A with H, V axes rotation and X,Y,Z linear movement
DUT positioning in X,Y axes on a motorized table
All axes equipped with absolute encoders
Angular range H axis± 180°
Nominal torque 98 Nm. Speed up to 50°/s
Angular resolution of encoder reading for H axis0.002°
Angular positioning accuracy for H axis0.05° *
Angular range V axis± 100°
Nominal torque 555 Nm. Speed up to 10°/s
Angular resolution of encoder reading for V axis0.002°
Angular positioning accuracy for V axis0.05° *
Linear range for X axis± 300 mm
Speed up to 40 mm/s
Linear range for Y axis± 150 mm
Speed up to 40 mm/s
Linear range for Z axis0-600 mm
Lifting capability up to 1500 N
Linear positioning accuracy for X,Y,Z70 μm
DUT photometric center positioningX,Y,Z linear movement, motorized
Photometric distance100 ft (FMVSS) 25 m, 10 m, 3,162 m (ECE) or other , user defined
DUT mounting tableSquare 500x500 mm. T-slot array 5x5 using Bosch Rexroth system, Multiple M6 mounting holes
Maximum DUT dimensions≤ 1800 mm (DUT centered)
Goniometer dimensions800 x 1780 x 2120 mm (W x H x D)
Control panel dimensions500 x 1230 x 500 mm (W x H x D)
Optical axis height1500 mm
Required spaceSquare 3000 x 3000 mm
Maximum load50 kg
Goniometer weight600 kg
Controls▪ GL GONIOMETER software for PC
▪ Remote manufacturer support capability
▪ 10 inch LCD touch display on the external HMI contr ol panel
▪ Manual controller for operations during DUT positioning
▪ Optional 10 channel multiplexer unit
CommunicationEthernet
Safety▪ External control panel with emergency stop button
▪ w/ customizable safety area, max. 3 m radius, 270°
Connections▪ a pair of banana sockets for DUT power supply, rated at 10 A max.
▪ a pair of banana sockets for remote sensing
▪ a single banana socket for protective connection
▪ 3.5 mm TRS jack for temperature probes
All sockets appear on the base of the goniometer as well as on the DUT mounting table
(or the optional multiplexer unit)
Power supply and max. ConsumptionAC 3-phase 400 V, 50 Hz, 3000 W **
Sensor typeGL OPTIC measuring instruments:
▪ GL SPECTIS 1.3 LS
▪ GL SPECTIS 1.0 Touch + Flicker
▪ GL SPECTIS 4.0 M
▪ GL SPECTIS 5.0 Touch
▪ GL PHOTOMETER 3.0 LS + Flicker
▪ GL SC RADIOMETER
▪ GL RETROREFLECTOMETER 4.0 SRS
Product no.202522
* at nominal load
** for the European market
Note: Instrument, firmware and software specification are subject to change without prior notice. All information included in GL OPTIC datasheets and product information available in any form are carefully prepared and included information believed to be true. Please note that discrepancies may occur due to text and/or other errors or changes in the available technology. We advise to contact GL Optic before the use of the product to obtain the latest product specification.

Videos

Related products

GL PHOTOMETER 3.0 + FLICKER

GL PHOTOMETER 3.0 + Flicker is a high sensitivity and high sampling rate photometer capable of extensive measurements of light flicker in addition to regular photometry.

GL SPECTROSOFT M

GL SPECTROSOFT M is an analytical PC-based software designed for laboratory applications, field work, production quality control and for general light assessment purposes.

GL SPECTIS 4.0 M

Precise testing of special UV VIS NIR lamps and sources.

Get quote

This post is also available in de_DE and pl_PL.